Detail AEMC AC Current Probe Model SR759
Alat Laboratorium Mikrobiologi
Deskripsi produk
AC Current Probe Model SR759
The SR759 probe is a range selectable probe that offers four switch selectable ranges: 1, 10, 100 or 1000A. This probe has a voltage output proportional to the current measured in the jaw of the probe. The SR759 has an integral 5 ft (1.5m) lead terminated by shrouded banana plugs compatible with DMMs and other instruments. The high accuracy, high sensitivity and low phase shift makes this probe an excellent choice for use in a wide range of power quality applications, including very low or leakage current measurements.
Specifications :
Maximum Conductor Size
2.05" (52mm)
Current Range
1A, 10A, 100A, 1000Arms AC
Measurement Range
1mA to 1.2A, 0.01 to 12A, 0.1 to 120A, 1 to 1200A
Output
5 ft (1.5m) lead with safety 4mm banana plug
Transformation Ratio
1:1, 10:1, 100:1, 1000:1
Output Signal
1000mV/AAC (1V @ 1A), 100mV/AAC (1V @ 10A), 10mV/AAC (1V @ 100A), 1 mV/AAC (1V @ 1000A)
Accuracy 50/60Hz
0.7% ± 1mV @ 1A, 0.5% @ 10A, 0.2% @100A, 0.2% @ 1000A
Frequency Range
30Hz to 3kHz with current derating
Phase Shift at 50Hz
1° @ 1000A
Open Secondary Voltage
8V peak
Load Impedance
10MΩ, 100kΩ min with derating
Overload
1200A for 40 mn ON then 20 mn OFF
Dimensions (H x W x D)
4.37 x 8.50 x 1.77” (111 x 216 x 45mm)
Weight
1.21 lbs (550g)
Operating Temperature
14° to +122°F (-10° to +50°C)
Storage Temperature
-4° to +158°F (-20° to +70°C)
Operating Humidity
90% RH
Pollution Degree
2
Ingress Protection (Ip)
IP40
EMC
EN 61326-1
Safety Compliance
EN/IEC 61010-1 600V CAT III, EN/IEC 61010-032