Detail AEMC AC/DC Current Probe Model MR525 - (Replaces Model MR520 - Cat.#12
Alat Laboratorium Mikrobiologi
Deskripsi produk
AC/DC Current Probe Model MR525 - (Replaces Model MR520 - Cat.#1200.7)
The MR525 is a hall effect current probe with a single range designed for DMM, Data loggers and Power Analyzers. The probe has a 5 ft (1.5m) lead terminated by 4mm safety banana plugs and can measure up to 1000AAC and 1400ADC. The probe is powered by a 9V battery or by a standard 5V external power supply via a micro-USB connector (not supplied).
The instrument's design includes internal temperature compensation.
Specifications :
Maximum Conductor Size
Cables: One 1.5” (39mm) or two 1” (25.4mm)
Bus Bar: One 1.97 x .49” (50 x 12.5mm) or two 0.98 x 0.2” (25 x 5mm); 1.24” x .30” (31.5 x 10mm) or three 0.98” x 0.31” (25 x 8mm)
Current Range
0.5 to 1000AAC, 1400ADC,
Output Signal
1mV/A via a 5 ft (1.5m) lead with 4 mm safety banana plugs
Accuracy 50/60Hz
≤2.5% @ 800A
Frequency Range
DC to 30kHz (-3dB) (depending of current value)
Phase Shift At 50Hz
≤1.5°@ 800A
Load Impedance
>1MΩ and ≤100pF
Overload
3000ADC or 1000AAC continuous for < 1kHz
Zero Adjust
Automatic on both ranges
Power Supply
9V alkaline Battery (NEDA 1060A, 6LR61) or 5V DC Micro-USB type B
Battery Life
50H typical
Low Battery Indication
Green LED blinking
Overload Indication
Red LED on when measurement > than selected range
Dimensions (H x W x D)
9.31 x 3.82 x 1.73” (236.5 x 97 x 44mm)
Weight (with Battery)
1.15lbs (520g)
Operating Temperature
14° to +131°F (-10 to +55°C)
Storage Temperature
-40° to +176°F (-40 to +80°C)
Relative Humidity
Up to 85% @ 35°C
Ingress Protection (Ip)
IP 40
EMC
EN 61326-1
Safety Compliance
IEC 61010-1, EN 61010-2-32, Pollution Degree 2, 600V CAT III