Detail AM1 Alcohol Analyser
AM1 Alcohol Analyser
The AM1 analyser is used for Clinical & Research applications and offers researchers into clinical alcoholism a very fast method of obtaining results with small animal models. In addition the small sample size ( 5ul) enables real-time results to be obtained during experiments.
APPLICATION AREAS
Alcoholism research
OPERATION
After calibration with a standard of the selected metabolite, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.
Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds
INSTRUMENT SPECIFICATION
Method
>
Enzymatic oxygen-rate
Sensor
>
Clark-type amperometric oxygen electrode
Reaction Temp.
>
30° C
Display
>
32 character backlit LCD
Printer
>
16 column dot matrix, 1 line/ sec.
Statistics
>
Sequential, giving mean, S.D and C.V.
Interface
>
Serial data port, optional Windows software available
Power
>
100-250VAC, 50-60Hz, 12-15VDC, 60VA
Dimensions
>
23cm( width) x 29cm( depth) x 15cm( height)
Weight
>
3.8Kg ( Portable, 6Kg)
Notes:
N/ A
ANALYTICAL PERFORMANCE
Precision
Accuracy ( Method Comparisons)
Linearity
Ethanol ( Alcohol)
>
C.V. of 2.5%
@ 18.5 mmol/ L
( ca. 85 mg/ dl)
( whole blood)
i) Method comparison for whole blood ( neutralised/ PCA extract) vs GC:
y ( Analox) = 1.039x + 0.28 mmol/ L,
r = 0.991, n = 27
ii) Urine Recovery Data:
y ( Analox) = 0.981x + 0.19 mmol/ L,
r = 0.999, n = 17
43.0 mmol/ L ( ca. 200 mg/ dl) for 5 µ l samples;
86.0 mmol/ L ( ca. 400 mg/ dl) for 2.5 µ l samples
Tampilkan Lebih Banyak