Back to Top

1
Hi, Guest!
  • PCE PCE
  • Clamp-on Ultrasonic Flow Meter PCE-TDS 200+ MR incl. ISO Calibration C

Clamp-on Ultrasonic Flow Meter PCE-TDS 200+ MR Incl. ISO Calibration C

Update Terakhir
:
28 / 08 / 2024
Min. Pembelian
:
1 Unit

Harga

Rp. 100
Contact :
Marketing : Bpk Juni / 08111980702
Marketing : Ibu Pera / 085210207070
Telp : 021 - 42803207
Fax : 021 - 42803208
email : [email protected]

Suplier alat-alat laboratorium & Intrument untuk berbagai keperluan alat-alat laboratorium Microbiology, Chemical and Reagent, Equipment, Medical dan Glassware.

Detail Clamp-on Ultrasonic Flow Meter PCE-TDS 200+ MR Incl. ISO Calibration C

Instrument Laboratorium

Deskripsi produk
DESCRIPTION
Ultrasonic flowmeter with 32 GB data memory / measuring range ±32 m/s /
reproducibility ±0.5 % of measured value / different measuring probes / alarm limit values /
2.8 " LC display / for all media

The ultrasonic flowmeter has a measuring range of ±32 m/s. With an accuracy of ±1.5 % f.s. for a pipe diameter of DN ≥50, ±3.5 % f.s. for a pipe diameter of DN <50 and a reproducibility of ±0.5 % f.s., the Clamp-on Ultrasonic Flow Meter is a particularly precise measuring device. An installation aid is available for installing the sensors of the Clamp-on Ultrasonic Flow Meter. The installation aid graphically displays the signal quality of the Clamp-on Ultrasonic Flow Meter. In addition, it is graphically displayed whether the sensors of the Clamp-on Ultrasonic Flow Meter are positioned at the correct distance from each other. To carry out flow measurements with the Clamp-on Ultrasonic Flow Meter, the flow velocity, the volume flow and the volume are displayed after entering the pipe and medium specifications. The Clamp-on Ultrasonic Flow Meter displays the measured values in a wide range of units. For example: m³, l, gal, igl, mgl, cf, bal, ib and ob.

During a measurement, it is possible to record the measured values via the data logger function of the Clamp-on Ultrasonic Flow Meter. Start and stop conditions and the storage interval can be set from 1 second to 12 hours for the Clamp-on Ultrasonic Flow Meter. A memory point of the Clamp-on Ultrasonic Flow Meter contains all measured values once. The measured values are stored on the 32 GB built-in memory of the Clamp-on Ultrasonic Flow Meter. 10 million measuring points can be stored on the Clamp-on Ultrasonic Flow Meter.

With the optional software, the recorded measured values can be read out from the Clamp-on Ultrasonic Flow Meter. For analysis, the measured values from the Clamp-on Ultrasonic Flow Meter can be displayed in tabular and graphic form. For further processing of the measured values from the Clamp-on Ultrasonic Flow Meter, they can be exported in .csv file format. Alternatively, a PDF report can be created via the software. For simplified operation, the Clamp-on Ultrasonic Flow Meter can be set via the software. A live view of the measured values from the Clamp-on Ultrasonic Flow Meter is also possible with the software.

In order to determine the amount of heat with the Clamp-on Ultrasonic Flow Meter, two additional thermocouples are required. These are available with the PCE-TDS 200+. The two thermocouples of the Clamp-on Ultrasonic Flow Meter are connected to the flow and return of a pipe system. Based on the temperature difference and the measured flow rate, the Clamp-on Ultrasonic Flow Meter can determine the amount of heat. If required, the Clamp-on Ultrasonic Flow Meter can calculate and display the costs per heat quantity unit simultaneously during the measurement. Thus, the Clamp-on Ultrasonic Flow Meter is used, for example, in the inspection of heating systems.

The LC colour display of the Clamp-on Ultrasonic Flow Meter has a size of 2.8" and is therefore easy to read.

PCE-TDS 200 MR SENSOR
Separate on-rail sensors for pipe sizes DN 50 ... 700 / 57 ... 720 mm. The temperature resistance is between -30 ... 160 °C. The sensor on the respective rail is located on a rail and can therefore be positioned appropriately. Magnets are located underneath the rails, which make it possible to hold on to pipes with a ferrous metal by themselves.
HIGHLIGHTS
- Measuring range ±32 m/s
- USB-C interface for data transfer
- Optional software for analysing the measured values
- Reproducibility ±0.5 % of measured value
- Heat quantity measurement
- Data memory for 10 million measuring points
- individually adjustable alarm limits
SPECIFICATION


Flow measurement

Measuring range
±32 m/s
Accuracy
0,001 m/s
Genauigkeit
DN ≥50 mm: ±1.5 % rdg. for velocities > 0.3 m/s

DN <50 mm: ±3.5 % rdg. for velocities > 0.3 m/s
Reproducibility
±0.5 % of measured value


PCE-TDS 200 MR SENSOR
Sensor pair
Pipe diameter
DN 50 … 700 / 57 … 720 mm
Temperature resistance
-30 ... 160 °C


Measuring method
Z, V, N, W
Medium
- water

- sea water

- oil

- crude oil

- methanol

- ethanol

- diesel

- petrol

- petroleum

- user defined

(manual input of the sound velocity from the medium)


All liquids with an impurity
<5 %
Pipe material
- copper CU

- steel FE

- stainless steel VA

- aluminium AL

- brass ME

- cast iron CI

- iron FE

- nickel NI

- titanium TI

- zink ZI

- acrylic AC

- polyethylene PE

- polypropylene PP

- polyvinyl chloride PVC

- nylon NY

- user defined

(manual input of the transversal sound velocity

of the pipe material)


Inner pipe lining
- no lining

- user defined

- epoxy resin

- rubber

- mortar

- polystyrene PS

- polyethylene PE

- polytetrafluoroethylene PTFE

- polyurethane PU

- polypropylene PP

- user defined

(manual input of the longitudinal sound velocity

of the inner lining of the pipe)


Measurement parameters
flow velocity, volume flow and volume
Units (dimensions)
mm, in
Units (flow velocity)
m/s, ft/s
Units (volume flow)
m³, l, gal, igl, mgl, cf, bal, ib, ob
Time specification
seconds, minutes, hours, days
Units (volume)
m³, l, gal, igl, mgl, cf, bal, ib, ob


Temperature measurement

Measuring range
Accuracy
Type B: 600 ... 1800 °C
Type B: ±(0.5 % + 3 °C)
Type E: -100 ... +900 °C
Type E: ±(0.4 % + 1 °C)
Type J: -100 ... 1150 °C
Type J: ±(0.4 % + 1 °C)
Type K: -100 ... +1370 °C
Type K: ±(0.4 % + 1 °C)
Type N: -100 ... + 1150 °C
Type N: ±(0.4 % + 1 °C)
Type R: 0 ... 1700 °C
Type R: ±(0.5 % + 3 °C)
Type S: 0 ... 1500 °C
Type S: ±(0.5 % + 3 °C)
Type T: -100 ... +400 °C
Type T: ±(0.4 % + 1 °C)

Tampilkan Lebih Banyak
Kontak Kami