Detail CO2 Analyser PCE-WMM 100
CO2 Analyser PCE-WMM 100
Instrument Laboratorium
Deskripsi produk
DESCRIPTION
CO2 Analyser PCE-WMM 100
CO2 warning device for carbon dioxide and oxygen / Adjustable alarms / 4-digit LC display / Consisting of main and remote unit / Relay contacts / Analog output
The CO2 warning device PCE-WMM 100 consists of a main unit and a remote display. The CO2 gas detector PCE-WMM 100 is used to measure carbon dioxide and oxygen in a fixed location. The CO2 warning device protects people from the odourless gas CO2 and also warns of low oxygen content in the air. The gas warning system displays the measured values on an LCD. The PCE-WMM 100 CO2 warning device has two relays that can be used to control alarm lights or sounders. Furthermore, the CO2 warning device has an integrated, acoustic alarm, which emits an audio signal of 80 db in the event of an alarm.
The warning device is used wherever CO2 can be released. This can be in warehouses, breweries, wine cellars or restaurants. The main unit is protected to IP54. 4 LEDs on the front provide information about the system status.
HIGHLIGHTS
- CO2 and O2 sensor
- Main and remote unit
- Alarm relay
- 4 ... 20 mA analog output
- Acoustic alarm
- Integrated temperature measurement
SPECIFICATION
Measuring function
CO2
Measuring range
0 .... 50000 ppm
Resolution
<10000 ppm: 10 ppm
>10001 ppm: 100 ppm
Accuracy
± 100 ppm / 5 % of measured value
Measuring function
O2
Measuring range
O2: 0 ... 30 %
Resolution
0.1 %
Accuracy
< 3 % of measured value
Measuring function
Temperature
Measuring range
0 ... 50 °C / 32 ... 122 °F
Resolution
0.1 °C
Accuracy
± 1 °C
Repeatability
CO2: ±20 ppm at 400 ppm
O2: < ±1 %
Temperature dependence
± 0.3 % of measured value / °C
(at 25 °C)
± 2 ppm / °C
(the larger value applies)
Pressure dependence
0.13 % of measured value / mmHg
Response time
CO2: T90 < 60s
O2: T90 < 30s
Temperature: 20 ... 30 m
Alarm 1
CO2: 5000 ppm, 1 %, 1.5 %, 2 %
O2: 18 %, 18.5 %, 19 %, 19.5 %, 20 %
Alarm 2
CO2: 1.5 %, 2 %, 2.5 %, 3 %, 3.5 %, 4 %
O2: 16 %, 16.5 %, 17 %, 17.5 %, 18 %
Alarm sound volume
80 db / 0.1 m
Heating time
< 60 s at 22 °C ambient temperature
Power supply
Power adapter 100 ... 240 V AC / 6 V DC / 0.5 A
Analog output
2 x 4 ... 20 mA
Load
<150 Ohm
Relay
max. 2 A / 30 V DC / 250 V AC
Operating conditions
0 ... 40 °C / 32 ... 104 °F / max 95 % r.H.
Storage conditions
-20 ... 60 °C / -4 ... 140 °F / max. 95 % r.H.
Degree of protection
Main unit: IP54
Remote unit: -
Dimensions
Main unit: 170 x 63 x 26 mm / 6.6 x 2.4 x 1.0"
Remote unit: 80 x 35.7 x 10 mm / 3.1 x 1.4 x 0.3"
Weight
Main unit: 1200 g incl cable
Remote unit: 120 g / 4.2 oz
DELIVERY SCOPE
1 x CO2 Analyser PCE-WMM 100
1 x Remote unit
1 x 8m connection cable
1 x Power adapter
1 x Mounting kit
1 x User manual
Tampilkan Lebih Banyak