Detail Current Clamp PCE-CM 3
Current Clamp PCE-CM 3
Instrument Laboratorium
Deskripsi produk
DESCRIPTION
Current clamp PCE-CM 3
Fork Clamp Up to 200 A AC / AC / DC Voltage Measurement / Simple Current Measurement / LC display / Continuity test / Thermometer
The current clamp is used for quick and easy AC measurements. For current measurement, the current clamp is placed over the current-carrying conductor. The current clamp is particularly suitable for the measurement of AC in distributions and wherever circuits must not be interrupted. The current measuring range extends from 0 ... 200 A. In addition to the current measurement, the current clamp can be used to measure DC and AC voltages up to 600V, resistances, capacitances and even temperatures. The measured values are displayed on the illuminated display of the current clamp.
The compact dimensions and the low weight distinguish this current clamp as well as the robust housing. This makes the current clamps the perfect companion for installers and service technicians.
HIGHLIGHTS
- Digital Multimeters up to 200 A
- Compact dimensions
- Robust plastic housing
- Multimeter functions
- Battery operation
- Backlit LCD
SPECIFICATION
DC
Measuring range
Resolution
Accuracy
4V DC
1 mV
± (1.2% of measured value + 2 digits)
40V DC
10 mV
± (1.5% of measured value + 2 digits)
400V DC
100 mV
± (1.5% of measured value + 2 digits)
600V DC
1V
± (2% of measured value + 2 digits)
Input impedance
10 MΩ
Overvoltage protection
600V DC
600V AC RMS
AC
Measuring range
Resolution
Accuracy
4V AC
1 mV
± (1.5% of measured value + 5 digits)
40V AC
10 mV
± (1.5% of measured value + 2 digits)
400V AC
100 mV
± (1.5% of measured value + 2 digits)
600V AC
1 V
± (2% of measured value + 2 digits)
Input impedance
10 MΩ
Overvoltage protection
600V DC
600V AC RMS
Frequency range
50 ... 400 Hz
Alternating current
Measuring range
Resolution
Accuracy
200 A AC
100-mA
± (3.0% of measured value + 5 digits)
Overload protection
200 A AC
Frequency range
50 ... 60 Hz
Resistance
Measuring range
Resolution
Accuracy
400 Ω
0.1 Ω
± (1% of measured value + 4 digits)
4 kΩ
1 Ω
± (1.5% of measured value + 4 digits)
40 kΩ
10 Ω
± (1.5% of measured value + 4 digits)
400 kΩ
100 Ω
± (1.5% of measured value + 4 digits)
4 MΩ
1 kΩ
± (2.5% of measured value + 4 digits)
40 MΩ
10 kΩ
± (3.5% of measured value + 4 digits)
Overvoltage protection
250V DC
250 AC RMS
Connection thermocouple
Type K connection with adapter
Capacity
Measuring range
Resolution
Accuracy
4 nF
0.1 nF
± (5% of measured value + 20 digits)
40 nF
1 nF
± (3% of measured value + 5 digits)
400 nF
10 nF
± (3% of measured value + 5 digits)
4 μF
100 nF
± (3% of measured value + 5 digits)
40 μF
1 μF
± (3% of measured value + 5 digits)
100 μF
10 μF
± (3% of measured value + 10 digits)
Overvoltage protection
250V DC
250 AC RMS
Check feature
Test
Display
Diode test
Test current: 0.5-mA
Reverse Voltage: 1.5V
Forward voltage of the diode
Continuity test
Open circuit Voltage 0.5V
Noise when resistance <50 Ω
Overvoltage protection
250V DC
250 AC RMS
General technical specifications
Jaw Capacity
About 17 mm / .7 in
Display
4000 digit LCD with backlight
Continuity test
Noise when resistance <50 Ω
Test current
About 0.5-mA
Open circuit voltage
<2V DC
Battery indicator
Battery icon when battery voltage low
Display overrange
OL appears in the display
Refresh rate display
3 Hz
Temperature sensor
Thermocouple type K
Input impedance
10 MΩ
Operating conditions
5 ... 40°C / 41 ... 104°F, 80% rh non-condensing
Storage conditions
-20 ... 60°C / -4 ... 140°F, 80% rh non-condensing
Operating altitude
<2000 m
Power supply
2 x 1.5V AAA battery
Automatic shutdown
30 minutes
Security conditions
IEC1010-1 (2001)
CAT II 1000V
CAT III 600V
Pollution degree
2
DELIVERY SCOPE
1 x Current clamp PCE-CM 3
2 x Test leads
2 x Measuring tip
1 x Adapter thermocouple
1 x Thermocouple type K
2 x AAA batteries 1.5V
1 x User manual
Tampilkan Lebih Banyak