Back to Top

1
Hi, Guest!
  • PCE PCE
  • Current Clamp PCE-PA 8000

Current Clamp PCE-PA 8000

Update Terakhir
:
04 / 09 / 2024
Min. Pembelian
:
1 Unit

Harga

Rp. 100
Contact :
Marketing : Bpk Juni / 08111980702
Marketing : Ibu Pera / 085210207070
Telp : 021 - 42803207
Fax : 021 - 42803208
email : [email protected]

Suplier alat-alat laboratorium & Intrument untuk berbagai keperluan alat-alat laboratorium Microbiology, Chemical and Reagent, Equipment, Medical dan Glassware.

Detail Current Clamp PCE-PA 8000


Current Clamp PCE-PA 8000
Instrument Laboratorium

Deskripsi produk
DESCRIPTION
PCE-PA 8000 Current Clamp
real time measurement with registration in the SD card / Integrated timer and calendar to record data / Active, apparent, and reactive power

The PCE-PA 8000 Current Clamp is able to measure and record the power in both single-phase or three-phase circuits. With the PCE-PA 8000 Current Clamp you will be able to perform prolonged measurements. Measurement values are saved in xls format on the included SD card of our Current Clamp to be easily analyzed on the computer. Measurement sample rate can be set from 2 to 7200 seconds. The PCE-PA 8000 Current Clamp can also measure power and determine both power factor and phase angle. The clamps of the PCE-PA 8000 Current Clamp can be used for cables with a diameter of 50 mm / 2.0 in. The PCE-PA 8000 is the ideal tool to be used in power supply networks. Its 3.7 in / 94 mm display allows you to read all measured values for a quick analysis as well as providing a view of the current, voltage, and power.

HIGHLIGHTS
- Analysis of a network of 3 phases / 4 conductors, 3 phases / 3 conductors, 1 phase / 2 conductors, 1 phase / 3 conductors
- Measures effective real value of current and voltage
- Measures active, apparent, and reactive power
- Power Factor (PF), phase angle (Φ)
- Measures voltage up to 600 V
- Measures current up to 1200 A
- It meets the IEC 1010 and CAT III 600 V standards
- Integrated timer and calendar to record data
- Measured values are recorded in the included SD card
- Data is stored in Excel format for further study
- Data can be transferred to a PC through data cable
- Large 3.7 in / 94 mm LCD display

SPECIFICATION
Technical Specifications
Voltage (AC V)
Measurement range
Resolution
Accuracy
10 V ... 600 V (phase to neutral)
0.1 V
±(0.5 % + 0.5 V)
10 V ... 600 V (phase to neutral)
Current (AC A)
Measurement range
Resolution
Accuracy
20 A
0.001 A (< 10 A)
0.01 A (≥ 10 A)
±(0.5 % + 0.1 V)
200 A
0.01 A (< 100 A)
0.1 A (≥ 100 A)
±(0.5 % + 0.5 A)
1200 A
0.1 A (< 1000 A)
1 A (≥ 1000 A)
±(0.5 A + 5 A)
Active power
Measurement range
Resolution
Accuracy
0 KW ... 9.999 KW
0.001 KW
±(1 % + 0.008 KW)
10 KW ... 99.99 KW
0.01 KW
±(1 % + 0.008 KW)
100 KW ... 999.9 KW
0.1 KW
±(1 % + 0.8 KW)
1 MW ... 9.999 MW
0.001 MW
±(1 % + 0.008 MW)
Apparent power
Measurement range
Resolution
Accuracy
0 KVA ... 9.999 KVA
0.001 KVA
±(1 % + 0.008 KVA)
10 KVA ... 99.99 KVA
0.01 KVA
±(1 % + 0.08 KVA)
100 KVA ... 999.9 KVA
0.1 KVA
±(1 % + 0.8 KVA)
1 MVA ... 9.999 MVA
0.001 MVA
±(1 % + 0.008 MVA)
Reactive power
Measurement range
Resolution
Accuracy
0 KVAR ... 9.999 KVAR
0.001 KVAR
±(1 % + 0.008 KVAR)
10 KVAR ... 99.99 KVAR
0.01 KVAR
±(1 % + 0.08 KVAR)
100 KVAR ... 999.9 KVAR
0.1 KVAR
±(1 % + 0.8 KVAR)
1 MVAR ... 9.999 MVAR
0.001 MVAR
±(1 % + 0.008 MVAR)
Active energy
Measurement range
Resolution
Accuracy
0 KWh ... 9.999 KWh
0.001 KWh
±(1 % + 0.008 KWh)
10 KWh ... 99.99 KWh
0.01 KWh
±(1 % + 0.08 KWh)
100 KWh ... 999.9 KWh
0.1 KWh
±(1 % + 0.8 KWh)
1 MWh... 9.999 MWh
0.001 MWh
±(1 % + 0.008 MWh)
Apparent energy
Measurement range
Resolution
Accuracy
0 KVAh ... 9.999 KVAh
0.001 KVAh
±(2 % + 0.008 KVAh)
10 KVAh ... 99.99 KVAh
0.01 KVAh
±(2 % + 0.08 KVAh)
100 KVAh ... 999.9 KVAh
0.1 KVAh
±(2 % + 0.8 KVAh)
1 MVAh... 9.999 MVAh
0.001 MVAh
±(2 % + 0.008 MVAh)
Reactive energy
Measurement range
Resolution
Accuracy
0 KVARh ... 9.999 KVARh
0.001 KVARh
±(2 % + 0.008 KVARh)
10 KVARh ... 99.99 KVARh
0.01 KVARh
±(2 % + 0.08 KVARh)
100 KVARh ... 999.9 KVARh
0.1 KVARh
±(2 % + 0.8 KVARh)
1 MVARh... 9.999 MVARh
0.001 MVARh
±(2 % + 0.008 MVARh)
Power factor
Measurement range
Resolution
Accuracy
0 ... 1
0.01
±0.04
Phase angle
Measurement range
Resolution
Accuracy
-180º ... 180º
0.1º
±1º
Frequency
Measurement range
Resolution
Accuracy
45 Hz ... 65 Hz
0.1 Hz
0.1 Hz
General specifications
Display
3.7 in / 94.0 mm LCD (320 x 240 pixels) with LED backlight
Electrical properties
AC V, AC A, active power, apparent power, power factor, phase angle, mains frequency
Voltage ranges
10 V ... 600 V AC, automatic range selection
Security / Standards
IEC1010CAT III 600V
AC V input resistance
10 MΩ
Operation frequency of current clamps
40 Hz ... 1 kHz
Tested operation frequency of current clamps
45 Hz ... 65 Hz
Overload protection
AC V / 720 V rms
AC A
1300 A with current clips
Memory
SD card up to 16 GB
Measurement display rate
1 second
Data logger
SD
Measurement sample rate
2 seconds... 7200 seconds
Data output
USB or RS-232, depending on the connection cable
Operating temperature
0º C ... 50º C / 32° F ... 122° F
Operating humidity
< 80 % H.r.
Power
8 x 1.5 V AA batteries / mains component
Input current
Meter: 300 mA DC
Clamp: 34 mA DC
Max. diameter
50 mm / 1.9 in
Weight
Meter: 948 g / 33.4 oz (with batteries)
Clip: 467 g / 16.4 oz (with cable)
Dimensions
Meter: 225 mm x 125 mm x 64 mm / 8.8 in x 4.9 in x 2.5 in
Clamp: 210 mm x 64 mm x 33 mm / 8.2 in x 2.5 in x 1.2 in
Jaw: 86 mm / 3.3 in (exterior)

DELIVERY SCOPE
1 x PCE-PA 8000 Current Clamp
3 x Current clamps
4 x Test cables
4 x Alligator clips
1 x 2 GB SD card
1 x 9 V DC mains component
8 x Batteries 1.5 V AA
1 x Carrying case
Tampilkan Lebih Banyak