Detail Current Clamp PCE-PCM 3
Current Clamp PCE-PCM 3
Instrument Laboratorium
Deskripsi produk
DESCRIPTION
Current clamp PCE-PCM 3
Digital multimeter for single-phase measurement / NCV measurement / W / VA / Var / Inrush Current / 1000 A AC Current Measurement / Bluetooth Interface / Frequency Measurement / Backlight / Voltage measurement / 750 kW power
With the current clamp AC currents of up to 1000 A AC can be measured easily. At the same time, voltages of up to 1000V AC / DC can be measured with the current clamp. Thus, with the power current clamp, apparent power, reactive power and active power measurements up to 750 KW are possible. Another function of the current clamp is the measurement of the inrush current. Many consumers have a tremendously high current at the moment of switch-on. With a conventional current clamp such a measurement is not possible. For the current clamp this is not a problem.
With the NCV (Not Connected Voltage) function of the current clamp, it can be checked whether a voltage is applied to a line without performing a contacting measurement. This makes the work much easier for the user thanks to the current clamp. In addition to the measuring function, the current clamp can also be connected to a terminal via the Bluetooth interface. Via this Bluetooth interface, the measured data can be transmitted directly from the current clamp to the terminal, where it can be logged, analyzed and sent.
HIGHLIGHTS
- Apparent / active / reactive power
- Measurement up to the 20th harmonic
- "Not Connected Voltage" function
- Inrush current measurement
- Current measurement up to 1000 A AC
- Bluetooth interface
SPECIFICATION
Ambient conditions 23°C / 73.4°F ± 5°C / 41°F, < 75% RH
Measuring function
Current measurement AC
Measuring range
Resolution
Accuracy
60 A
0.01 A
± 2% + 8 Dgt
600 A
0.1 A
± 2% + 8 Dgt
1000 A
1 A
± 2% + 8 Dgt
Maximum input current: 1000 A AC
Frequency range: 40 .... 400 Hz
Measuring function
Inrush current AC
Measuring range
Resolution
Accuracy
60 A
0.01 A
For reference only
600 A
0.1 A
± 5% + 60 Dgt
1000 A
1 A
± 5% + 60 Dgt
Integral time: 100 ms
Measuring range: 20 ... 1000 A
Frequency range: 40 .... 400 Hz
Measuring function
Voltage measurement DC
Measuring range
Resolution
Accuracy
6V
0.001V
± 0.5% + 5 Dgt
60V
0.01V
± 0.5% + 5 Dgt
600V
0.1V
± 0.5% + 5 Dgt
1000V
1V
± 0.8% + 4 Dgt
Input impedance: 10 MΩ
Maximum input voltage: 750V AC (rms value) or 1000V DC
Measuring function
Voltage measurement AC
Measuring range
Resolution
Accuracy
6V
0.001V
± 0.6% + 4 Dgt
60V
0.01V
± 0.6% + 4 Dgt
600V
0.1V
± 0.6% + 4 Dgt
1000V
1V
± 0.8% + 4 Dgt
Input impedance: 10 MΩ
Maximum input voltage: 750V AC (rms value) or 1000V DC
Frequency range: 40 ... 400 Hz
Measuring function
Frequency measurement with
The current measuring function
Measuring range
Resolution
Accuracy
99.99 Hz
0.01 Hz
± 1.5% + 5 Dgt
999.9 Hz
0.1 Hz
± 1.5% + 5 Dgt
Measuring range: 10 ... 1 kHz
Measuring range of the input signal: > 20 AAC (rms value)
The input current increases with the measuring frequency.
Maximum input current: 1000 AAC (rms value)
Measuring function
Frequency measurement with the
Voltage measurement function
Measuring range
Resolution
Accuracy
99.99 Hz
0.01 Hz
± 1.5% + 5 Dgt
999.9 Hz
0.1 Hz
± 1.5% + 5 Dgt
9,999 kHz
0.001 kHz
± 1.5% + 5 Dgt
Measuring range: 10 ... 10 kHz
Measuring range of the input signal: > 1V AC (rms value)
The input current increases with the measuring frequency.
Maximum input voltage: 1000V AC (rms value)
Measuring function
Frequency measurement
Measuring range
Resolution
Accuracy
9,999 Hz
0.001 Hz
± 0.3% + 5 Dgt
99.99 Hz
0.01 Hz
± 0.3% + 5 Dgt
9,999 kHz
0.1 Hz
± 0.3% + 5 Dgt
999.9 Hz
0.001 kHz
± 0.3% + 5 Dgt
99.99 kHz
0.01 kHz
± 0.3% + 5 Dgt
999.9 kHz
0.1 kHz
± 0.3% + 5 Dgt
9,999 MHz
0.001 MHz
± 0.3% + 5 Dgt
Tampilkan Lebih Banyak