  • Earth Tester / Earth Resistance Meter PCE-ET 5000

Earth Tester / Earth Resistance Meter PCE-ET 5000

Update Terakhir
:
02 / 09 / 2024
Min. Pembelian
:
1 Unit

Harga

Rp. 100
Contact :
Marketing : Bpk Juni / 08111980702
Marketing : Ibu Pera / 085210207070
Telp : 021 - 42803207
Fax : 021 - 42803208
email : [email protected]

Suplier alat-alat laboratorium & Intrument untuk berbagai keperluan alat-alat laboratorium Microbiology, Chemical and Reagent, Equipment, Medical dan Glassware.

Detail Earth Tester / Earth Resistance Meter PCE-ET 5000


Earth Tester / Earth Resistance Meter PCE-ET 5000
Instrument Laboratorium

Deskripsi produk
DESCRIPTION
Earth Tester / Earth Resistance Meter PCE-ET 5000
Grounding meter with memory / Mobile use possible thanks to battery / Resistance measurement / Protective conductor test / Visual connection aid / Lightning conductor test / Backlight Testing of disturbing voltages, frequencies and currents
The earth tester / earth resistance meter is a versatile ohmmeter. With this grounding meter a variety of resistance measurements can be performed. Thus, with the grounding meter, the possibility exists to perform a normal resistance measurement of 2- or 4-pole application. It is also possible to determine the resistance of the soil with the grounding meter. For the measurement of earth resistance corresponding earth spikes are included in the scope of supply. These ground spikes can easily be connected to the grounding device via the supplied separate 20 meter / 66 ft cable.

Another important measuring function is the measurement of the resistance of, for example, lightning conductors. This measurement is used to control the lightning conductors on, for example, schools or university buildings and is essential to derive a possible lightning strike into the ground. This earth tester / earth resistance meter can thus be used to determine whether the lightning rod is properly connected to the ground and whether it has the right resistance value to discharge a lightning bolt into the ground.

To ensure a perfect measurement, the earth tester / earth resistance meter has a measuring terminal detection. After the operator has set the grounding meter to the desired measurement function, the terminals to be used light up. If an incorrect connection has been selected by the operator, the grounding meter signals the operator both acoustically and visually. At the same time, the measurement by the grounding meter is refused until the test setup has been properly connected. This massively limits faulty measurements by the intelligent grounding meter. Also, the grounding device has a memory. This allows the measurement data from the grounding meter to be transferred to a PC and evaluated after a measurement.

HIGHLIGHTS
- Mobile use thanks to battery and batteries
- Data memory for later analysis on the PC
- Visual connection aid for easy operation
- Alarm in case of incorrect connection of the measuring line
- Direct and indirect resistance measurement
- Interference currents, interference voltages, frequencies
SPECIFICATION
3-pole earth resistance

Test voltage
20, 48V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
26 seconds
Probe resistance Rs
<100 kΩ
Auxiliary earth resistance Rh
<100 kΩ
Resolution
0.001 Ω
Measuring range Re
0.02 Ω ... 300 kΩ
Accuracy
± (5% + 10 digits)
Ust
< 24V
Fst
16 ... 400 Hz


The accuracy is ensured in an environment of 0 ... 28°C / 32 ... 82°F, < 80% RH (non-condensing).
Response time < 15 seconds (Time required to reach the specified accuracy, after the start of the measurement with the average function switched off.)


4-pole grounding resistor

Test voltage
20, 48V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
30 seconds
Probe resistance Rs
< 100 kΩ
Auxiliary earth resistance Rh
< 100 kΩ
Resolution
0.001 Ω
Measuring range Re
0.02 Ω ... 300 kΩ
Accuracy
± (5% + 10 digits)
Ust
< 24V
Fst
16 ... 400 Hz


The accuracy is ensured in an environment of 0 ... 28°C / 32 ... 82°F, < 80% RH (non-condensing).
Response time < 25 seconds (Time required to reach the specified accuracy, after the start of the measurement with the average function switched off.)


3-pole earth resistance with current clamp
Test voltage
20, 48V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
26 seconds
Probe resistance Rs
< 100 kΩ
Auxiliary earth resistance Rh
< 100 kΩ
Resolution
0.001 Ω
Measuring range Re
0.02 Ω ... 20 kΩ
Accuracy
± (5% + 10 digits)
Ust
< 24V
Fst
16 ... 400 Hz


The accuracy is ensured in an environment of 0 ... 28°C / 32 ... 82°F, < 80% RH (non-condensing).
Response time < 25 seconds (Time required to reach the specified accuracy, after the measurement has started with the average function switched off.)
If the current on the current probe is too low, the measurement can be aborted.


4-pole earth resistance with current clamp
Test voltage
20, 48V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
26 seconds
Probe resistance Rs
< 100 kΩ
Auxiliary earth resistance Rh
< 100 kΩ
Resolution
0.001 Ω
Measuring range Re
0.02 Ω ... 20 kΩ
Accuracy
± (5% + 10 digits)
Ust
< 24V
Fst
16 ... 400 Hz


The accuracy is ensured in an environment of 0 ... 28°C / 32 ... 82°F, < 80% RH (non-condensing).
Response time < 25 seconds (Time required to reach the specified accuracy, after the measurement has started with the average function switched off.)
If the current on the current probe is too low, the measurement can be aborted.


Grounding resistor only with current clamps
Test voltage
48V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
26 seconds
Measuring range Re
0.02 Ω ... 150 Ω
Resolution
0.001 Ω
Accuracy
± (5% + 10 digits)
Ust
< 24V
Fst
16 ... 400 Hz


The accuracy is ensured in an environment of 0 ... 28°C / 32 ... 82°F, < 80% RH (non-condensing).
Response time < 25 seconds (Time required to reach the specified accuracy, after the measurement has started with the average function switched off.)
If the current on the current probe is too low, the measurement can be aborted.


Soil resistance (removal resistance)
Test voltage
20, 48V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
26 seconds
Resolution
0.001 Ω * m
Measuring range Re
0.02 Ω ... 1000 kΩ * m
Accuracy
± (5% + 10 digits)


2-pole resistance measurement (AC)

Test voltage
20V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
26 seconds
Measuring range Re
0.02 Ω ... 300 kΩ
Resolution
0.001 Ω
Accuracy
± (5% + 10 digits)
Ust
< 24V
Fst
16 ... 400 Hz


2-pole resistance measurement (DC)

Test voltage
20V DC
Short circuit current
250-mA
Test time
26 seconds
Measuring range Re
0.02 ... 3 kΩ
Resolution
0.001 Ω
Accuracy
± (5% + 10 digits)
Ust
< 3V
Fst
16 ... 400 Hz


4-pole resistance measurement (DC)

Test voltage
20V DC
Short circuit current
250-mA
Test time
26 seconds
Measuring range Re
0.02 Ω ... 3 kΩ
Resolution
0.001 Ω
Accuracy
± (5% + 10 digits)
Ust
< 3V
Fst
16 ... 400 Hz


Line compensation

Test voltage
20, 48V AC
Test frequency
94 Hz, 105 Hz, 111 Hz, 128 Hz, AFC
Short circuit current
250-mA
Test time
26 seconds
Resolution
0.001 Ω
Measuring range Re
0.02 Ω ... 30 Ω
Accuracy
± (3% + 10 digits)
Ust
< 24V
Fst
16 ... 400 Hz


Interference voltage

Measuring range
± 50V DC

1 ... 50V AC
Resolution
0.1V
Accuracy
± (5% + 5 digits)


Interference current

Measuring range
20-mA ... 2 A
Resolution
1-mA
Accuracy
± (5% + 5 digits)