The HI-99301 is a portable, EC/TDS/temperature meter that measures high range conductivity in mS/cm and TDS in ppt. Both instruments are housed in a case rated for IP67 conditions.
This instrument offers single point, automatic calibration with automatically temperature compensated measurements. The compensation coefficient and EC/TDS conversion factor are user-selectable from 0.45 to 1.00 and temperature coefficient is selectable from 0.0 to 2.4%/°C.
thermometer
This meter easily fits in the palm of your hand and the bottom probe connection ensures the electrode cable doesn't get in your way. The large, multi-level LCD displays the primary reading, temperature and calibration guides simultaneously. Symbols and messages on the LCD indicate meter status and guides users through operations.
At start-up, the meter shows the remaining battery percentage and when a low battery condition is detected, a battery symbol appears on the LCD to advise the user that only a few hours of working time is left.
The HI-76306 EC/TDS/temperature probe resists clogging and is easy to clean.
Ideal for
Main benefits
- BEPS (Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings
- Battery % displayed on start-up
- Compact, heavy-duty and water resistant
Key features
- Amperometric probe technology
- Automatic Temperature Compensation
- Automatic single calibration with calibration indicator
- On-screen tutorial messages for calibration and set-up
Supplied with
- HI-76306 EC/TDS probe
- Batteries
- Instructions and rugged carrying case
- sPECIFICATION
- Range EC 0.00 to 20.00 mS/cm Range TDS 0.00 to 10.00 ppt (g/L) Range Temperature 0.0 to 60.0°C / 32.0 to 140.0°F Resolution EC 0.01 mS/cm Resolution TDS 0.01 ppt (g/L) Resolution Temperature 0.1°C / 0.1°F Accuracy EC ±2% F.S. Accuracy TDS ±2% F.S. Accuracy Temperature ±0.5°C / ±1°F TDS Factor adjustable, from 0.45 to 1.00 with 0.01 step
(default 0.50) Calibration automatic, 1 point at 6.44 ppt (CONV 0.5)
or at 9.02 ppt (CONV 0.7)
or at 12880 µS/cm (others) Temperature Compensation Automatic, 0 to 60°C (32 to 140°F) with ß adjustable
from 0.0 to 2.4% with 0.1% step Probe HI 76306, EC/TDS/temperature, with DIN connector,
1 m (3.3') cable (included) Battery Type / Life 4 x 1.5V AAA / approx. 500 hours of continuous use Environment 0 to 50°C (32 to 122°F); RH max 100% Dimensions 143 x 80 x 38 mm (5.6 x 3.1 x 1.5") Weight 210 g (7.4 oz.)