Detail Hioki Isplay Unit CM7290
Basic specifications (Accuracy guaranteed for 3 years, Post-adjustment accuracy guaranteed for 3 years)
Sensor
CT7642, 7742
CT7636, 7736
CT7631, 7731
Measurement parameters
DC, AC, DC+AC, Hz
Crest factor
3 at 5000 count or 2.5 at 6000 count for AC and DC+AC
Output method
WAVE, RMS, PEAK, FREQ
Input connectors
HIOKI PL 14
Output update time
PEAK--- FAST: 0.02 s / NORMAL: 0.2 s / SLOW: 1 s
FREQ--- FAST: FAST: 0.2 s / NORMAL: 0.2 s / SLOW: 3.0 s
(WAVE, RMS: analog output)
PEAK sensing duration
2 ms or greater
(during PEAK MAX/PEAK MIN and PEAK output)
Other functions
Auto range, Zero adjustment at power-up, Analysis display, Filter, Output amplification, Display value hold, Backlight , Auto-power save, Save settings, keypad lock
Typical accuracy (WAVE output DC)
±2.0% rdg. ±10.8 mV
(60.00 A range)
±2.5% rdg. ±30.8 mV
(60.00 A range)
±1.5% rdg. ±5.8 mV
(60.00 A range)
Typical accuracy (RMS output AC)
±2.3% rdg. ±10.8 mV
(60.00 A range)
±2.8% rdg. ±30.8 mV
(60.00 A range)
±1.8% rdg. ±5.8 mV
(60.00 A range)
Power supply
LR6 alkaline batteries (AA) ×2, Continuous use: 16 h (backlight OFF and WAVE or RMS output, when used with CT7600s), Rated power 2.5 VA
or AC adapter 9445-02/03 (100 to 240V AC), or 5 to 15 V DC external power supply, Rated power 2.5 VA
Dimensions and mass
52 mm (2.05 in)W × 163 mm (6.42 in)H × 37 mm (1.46 in)D, 220 g (7.8 oz) (including protector and battery)
Accessories
LR6 alkaline batteries ×2, Protector ×1, Instruction manual ×1
Tampilkan Lebih Banyak