HORIBA LAQUAact EC110
Alat Laboratorium Umum
Features
Automatic and manual conductivity calibration up to 4 points
Adjustable reference temperature and temperature coefficient
Selectable cell constant and measurement unit
Automatic / Manual temperature compensation and temperature calibration function
Conductivity and total dissolved solids (TDS) conversion
Four (4) TDS calibration curves for wide range of applications: EN27888, Linear, 442, NaCl
Built-in memory can store up to 100 data
Auto shut-off (programmable: 1 to 30 minutes)
Auto-Hold / Auto-Stable measurement
Large LCD: 52mm x 64mm
Powered by 2 x AAA batteries (LR03) or Ni-MH rechargeable batteries or power adaptor
Casing is IP67 waterproof (for 30mins immersion in 1m depth of water) and dustproof
Packaged in new molded carry case
2 years meter warranty
Specifications
Model
LAQUA PC1100
Dual Channel pH/ORP/Cond/TDS/Res/Sal/Temp
pH Range
-2.000 to 20.000 pH
Resolution
0.1, 0.01, 0.001 pH
Accuracy
± 0.003 pH
Calibration Points
Up to 5
Buffer Options
USA, NIST, Custom
ORP Range
± 1999.9 mV
Resolution
0.1 mV
Accuracy
± 0.2 mV
Conductivity Range
...µS/cm to 2000 mS/cm
Resolution
0.05% F.S.
Accuracy
±0.6% F.S.; ±1.5% F.S. > 18.0 mS/cm
Reference Temperature
15 to 30.0°C (adjustable)
Temperature Coefficient
0.0 to 10.0% (adjustable)
Cell Constant
0.1, 1.0, 10.0 (selectable)
Measurement Unit
μS/cm, mS/cm, S/m (selectable / auto-ranging)
Calibration Points
Up to 4 (Auto / Manual)
TDS Range
0.01 ppm to 1000 ppt
Resolution
0.01 ppm / 0.1 ppt
Accuracy
±0.1% F.S.
Calibration Curves
EN27888, 442, Linear, NaCl
Resistivity Range
0.000 Ω•cm to 20.000 MΩ•cm
0.00 Ω•cm to 200.0 MΩ•cm
Resolution
0.05% F.S.
Accuracy
±0.6% F.S.; ±1.5% F.S. > 18.0 MΩ•cm
Salinity Range
0.0 to 100.0 ppt / 0.00 to 10.00 %
Resolution
0.01 ppm / 0.1 ppt
Accuracy
0.2% F.S.
Calibration Curves
NaCl / Seawater
Temperature Range
-30.0 to 130.0°C
Resolution
0.1°C
Accuracy
± 0.4°C
Calibration Option
Yes
Memory Data
999
Auto Log Data
Yes
Real-time Clock
Yes
Date / Time Stamp
Yes
Auto-Shut Off
Yes (Programmable: 1 to 30 minutes)
Auto-Hold / Auto-Stable
Yes
Offset / Slope Display
Yes (Independent acid and alkaline slopes depending on calibration)
Electrode Status
On screen display
Diagnostic Messages
Yes
Calibration Alarm
Yes (Programmable: 1 to 400 days)
Display
Custom black and white static LCD, Dual Channel
Inputs
Dual BNC, dual phono, DC sockets
Output
RS232, USB
Power Requirement
AC adaptor, 100-240V, 50/60Hz
Weight
500g
Dimensions
170(L) x 174(D) x 73(H) mm