Detail KEW MATE 2000
DC V 340mV/ 3.4/ 34/ 340/ 600V ( Input Impedance: 10M© )
± 1.5% rdg± 4dgt
AC V 3.4/ 34/ 340/ 600V ( Input Impedance: 10M© )
± 1.5% rdg± 5dgt [ 50 - 400Hz]
DC A 60A ± 2% rdg± 5dgt
AC A 60A ± 2% rdg± 5dgt ( 50/ 60Hz)
© 340© / 3.4/ 34/ 340k© / 3.4/ 34M©
± 1% rdg± 3dgt ( 0 - 340k© )
± 5% rdg± 5dgt ( 3.4M© )
± 15% rdg± 5dgt ( 34M© )
Continuity buzzer Buzzer sounds below 30± 10©
( Continuity buzzer works on 340© range only)
Frequency ( AC A) 3.4/ 10kHz ± 0.1% rdg± 1dgt
( AC V) 3.4/ 34/ 300kHz ± 0.1% rdg± 1dgt
( Input sensitivity Current: more than 15A, Voltage: more than 30V)
Conductor size Æ 6mm max.
Withstand Voltage 3700V AC for 1 minute
Applicable standards IEC 61010-1 CAT III 300V Pollution degree 2
IEC 61010-031, IEC 61010-2-032, IEC 61326-1
Power Source R03 ( DC 1.5V) × 2
* Continuous measuring time: approx. 45 hours
( Auto power save: approx. 10 minutes)
Dimensions 128( L) × 87( W) × 24( D) mm
Weight 210g approx.
Accessories R03 ( DC 1.5V) × 2, Instruction Manual
Optional 9107 ( Carrying case [ Soft] )
Tampilkan Lebih Banyak