Detail KYORITSU AC/DC Digital Clamp Meters MODEL 2010
Alat Laboratorium Mikrobiologi
Deskripsi produk
• High sensitivity, miniature AC/DC clamp meter.
• 0.1mA minimum resolution for AC current and 1mA minimum resolution for DC current.
• Output terminal for recorder connection.
Never make current measurement on a circuit above 60V AC/DC
as the instrument is designed for use on low voltage circuits only.
AC A
200mA/2/20A
±1%rdg±2dgt [50/60Hz] (200mA)
±1.5%rdg±8dgt [40Hz - 2kHz] (200mA)
±1%rdg±2dgt [50/60Hz] (2A)
±2.5%rdg±10dgt [40Hz - 2kHz] (2/20A)
DC A
2/20A
±1%rdg±2dgt (2A)
±1.5%rdg±4dgt (20A)
Conductor size
Ø7.5mm max.
Frequency response
DC, 40Hz - 2kHz
Output
Recorder :
DC200mV against AC200mA/2/20A
DC200mV against DC2/20A
Power source
6LR61(9V) × 1 or AC Adaptor
Dimensions
142(L) × 64(W) × 26(D)mm
153(L) × 23(W) × 18(D)mm
Clamp Sensor
Weight
220g approx.
Included Accessories
9095 (Carrying Case)
Alkaline 6LF22 × 1
Instruction Manual
Optional Accessories
7256 (Output cord)