Detail KYORITSU Clamp Power Meter KEW 2060BT

Alat Laboratorium Mikrobiologi

Deskripsi produk
A tear drop shape jaw with the conductor size 55 mm: Convenient to use at a small-sized office and factory
・Current up to 1000A rms
・Voltage up to 1000V rms
・Harmonics up to 30th
・Various measurement functions: current, voltage, power, harmonics
and phase detection
・LCD can display simultaneously the values of voltage and frequency
or power and power factor
・Each degree of harmonics can be shown on the LCD of the tester
・Bluetooth® communication functions






1.Various measurement functions
〇 Measurements of current, voltage, power, harmonics and phase detection are possible.
〇 True RMS: Indicates a correct value at the distorted waveform.
〇 Functions of PEAK, MAX, MIN, AVG are available.
Easy to check the temporal change in the power source.
〇 LCD can display simultaneously the values of voltage and frequency or power and power
factor.



2.Power measurement
Power measurement on any wiring system is possible.



KEW 2062BT can perform Single-phase 2-Wire / Single-pahe 3-Wire
and balance and unbalance measurements of Three-phase 3-Wire
/ Three-phase 4 Wire. The double display can simultaneously
show many parameters like W & PF, W & deg, W & VA, W & Var,
V & A, etc.



Various parameters such as
active/reactive/apparent power,
power factor, phase differences
(1P2W only) which are required for the power measurement can be
measured.


Total 3-phase power can be
measured in simple steps.




By setting the CT and VT ratio and measuring the secondary side of
the transformation, the primary side value can be obtained.


3.Harmonics measurement
Harmonics of the voltage and current, which cause various failures at the field, can be measured.
It is possible to measure and show each voltage and current harmonics up to 30th.
Total harmonics distortion factor is in two types of THD-R and THD-F. Each degree of harmonics can be
shown on the LCD of the tester. Both the current value (or voltage value) and the content ratio can be displayed on the same screen.



4.Phase detection


5.Connectivity


○Dedicated Application "KEW Power*"supports both Android™ and iOS.
○LCD display can be checked at the smart devices.
○Measured values and graphs can be stored with a press of a button.
○Voltage and current can be shown in the waveform display. Existence of the harmonics can be found
easy.
○Threshhold of each measured value can be set at the applicaion, and PASS and FAIL judgement is
possible.

Wiring connections
1P2W, 1P3W, 3P3W, 3P4W
Measurements and
parameters
Voltage, Current, Frequency, Active power, Reactive power,
Apparent power, Power factor(cosθ), Phase angle,
Harmonics(THD-R/THD-F), Phase rotation
ACV

Range
1000V
Accuracy
±0.7%rdg±3dgt (40.0 - 70.0Hz)
±3.0%rdg±5dgt (70.1 - 1kHz)
Crest factor
1.7 or less
ACA

Range
40.00/400.0/1000A (3 range auto)
Accuracy
±1.0%rdg±3dgt (40.0 - 70.0Hz)
±2.0%rdg±5dgt (70.1 - 1kHz)
Crest factor
3 or less on 40.00A/400.0A range,
3 or less 1500A peak on 1000A range
Frequency

Display Range
40.0 - 999.9Hz
Accuracy
±0.3%rdg±3dgt
Active power

Range
40.00/400.0/1000kW
Accuracy
±1.7%rdg±5dgt (PF1, sine wave, 45-65Hz)
Apparent power

Range
40.00/400.0/1000kVA
Accuracy
±1dgt against each calculated value
Sum: add errors of each channel, 3P3W: ±2dgt, 3P4W:±3dgt
Reactive power

Range
40.00/400.0/1000kVar
Accuracy
±1dgt against each calculated value
Sum: add errors of each channel, 3P3W: ±2dgt, 3P4W:±3dgt
Power factor

Display Range
-1.000 - 0.000 - +1.000
Accuracy
±1dgt against each calculated value
Sum: add errors of each channel, 3P3W: ±2dgt, 3P4W:±3dgt
Phase angle (1P2W only)

Display Range
-180.0 - 0.0 - +179.9
Accuracy
±3.0°
Harmonics RMS(Content rate)

Analysis order
1st - 30th order
Accuracy
±5.0%rdg ± 10dgt (1 - 10th)
±10%rdg ± 10dgt (11 - 20th)
±20%rdg ± 10dgt (21 - 30th)
Total harmonics THD-R/THD-F

Display Range
0.0% - 100.0%
Accuracy
±1dgt against the calculated results of each measured value
Phase rotation
ACV 80 - 1100V (45 - 65Hz)
Other functions
MAX/MIN/AVG/PEAK, Data hold, BluetoothⓇ, Back light, Auto power off
General

Communication
interface
Bluetooth®5.0 Android™5.0 or more, iOS 10.0 or more
Power source
LR6(AAA)(1.5V) ×2
Continuous
measuring time
Approx. 58 hours
Conductor size
φ55mm max.
Dimensions / Weight
247(L) × 105(W) × 49(D) mm / Approx.490g(including batteries)
Applicable Standards
IEC 61010-1, IEC 61010-2-032, IEC 61326-1,-2-2 (EMC), IEC 60529(IP40)
CAT IV 300V / CAT lll 600V / CAT ll 1000V Pollution degree 2
Accessories
7290 (Voltage test lead set)
9198 (Carrying case)
LR6(AAA)×2, Instruction Manual
※ Some countries regulate the compliance with their Radio Law of the products equipped with BluetoothⓇ.
Please confirm it with your distributor before purchasing our products equipped with Bluetooth® .
※ BluetoothⓇ is a trademark or registered trademark of Bluetooth SIG. Inc. Android is a trademark or registered
trademark of Google Inc.
※ iOS is a trademark or registered trademark of Cisco Technology, Inc. in the United States and other countries.

Hubungi :

Juni Tarigan (Mr.)

0811 198 0702


Pera (Ms.)

0811 3801 0819


Admin

0813-2310-201


PT. ANDALAN BANGUN SEJAHTERA

Jl. Sumur Batu No. 1, Kemayoran

Jakarta Pusat - Indonesia 10640

Telp : +6221-42803207, 4227517


Fax : +6221-42803208
Tampilkan Lebih Banyak
Kontak Kami