Detail MEMMERT Universal Incubator
Ventilation and Control natural convection
continuous adjustment of pre-heated fresh air admixture
vent connection with restrictor flap
microprocessor PID-temperature controller with integrated autodiagnostic system with fault indicator
one Pt100 sensor class A in 4-wire-circuit
integrated digital timer ( 1 min. up to 99 h 59 min) to switch off heating to stand-by mode
digital LED- display of set and actual temperature as well as remaining process time
resolution of display for set value and actual value 0, 5 ° C
Overtemperature protection ( double)
in case of total breakdown of sensor, the heating is switched off at approx. 3 ° C above set value
in case of failure of the switching element, the heating will be controlled by an additional relay approx. 3 ° C above the set temperature
in case of a faulty sensor ( incorrect measuring values) or the total breakdown of the controller the heating is switched off at approx. 10 ° C above the max. oven temperature by the mechanical temperature limiter TB protection class 1
for very sensitive and/ or valuable load in terms of being non-reproducable, we recommend ovens of our E- and P-class guaranteeing an even higher safety standard
Textured Stainless Steel Casing
fully insulated stainless steel door with double locking and 4-point adjustment
inner glass door
rear zinc-plated steel
Interior - Heating Concept
easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
2 stainless steel grids
Temperature Range
from + 30 ° C ( however, at least 5 ° C above ambient) up to + 70 ° C
Voltage / Power Rating
230 V ( + / - 10% ) , 50/ 60 Hz
ca. 900 W ( during heating)
Results
Incubator IF30 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF30plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IF55 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF55plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IF75 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF75plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IF110plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IF110 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF160plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IF160 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF260 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF260plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IF450 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF450plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IF750 ( Forced air circulation ( Fan) / singleDISPLAY )
Incubator IF750plus ( Forced air circulation ( Fan) / twinDISPLAY )
Incubator IN30 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN30plus ( Natural air circulation ( Convection) / twinDISPLAY )
Incubator IN55 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN55plus ( Natural air circulation ( Convection) / twinDISPLAY )
Incubator IN75 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN75plus ( Natural air circulation ( Convection) / twinDISPLAY )
Incubator IN110 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN110plus ( Natural air circulation ( Convection) / twinDISPLAY )
Incubator IN160 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN160plus ( Natural air circulation ( Convection) / twinDISPLAY )
Incubator IN260 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN260plus ( Natural air circulation ( Convection) / twinDISPLAY )
Incubator IN450 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN450plus ( Natural air circulation ( Convection) / twinDISPLAY )
Incubator IN750 ( Natural air circulation ( Convection) / singleDISPLAY )
Incubator IN750plus ( Natural air circulation ( Convection) / twinDISPLAY )
Tampilkan Lebih Banyak